PROCESS/ON-LINE
Never before has a colorimeter offered as much versatility as
PSPI’s ClearView Analyzer. The 42015 CLV continuously mea-
sures color of liquid petroleum fuels, even in the presence of
haze. Easily determine color and haze using a single
measurement probe ...one sample, one test, two results! Sample
can be inserted directly in the pipeline or taken from a
side-stream. Various measurement probes, including high pres-
sure, high temperature and custom accommodate all your
testing needs. And because each insertion probe installs
directly into the pipeline, there is no need for a Fast Loop or
Sample Conditioning System. The analyzer is also available
with the Welker Insertion Device to assist with probe insertion
and retraction under pressure.
Measure both color and haze simultaneously,
continuously and completely on-line!
MEASUREMENT RANGE
ASTM D 1500 Scale
0 to 8
Saybolt (ASTM D 156) Scale
-16 to 30
APHA / Platinum Cobalt (ASTM D 1209)
0 to 100
Haze (ASTM D 4176)
1 to 6
Measures color on ASTM D 1500, Saybolt (ASTM D 156) or
APHA Platinum-Cobalt (ASTM D 1209) scales
Haze measurements correlate with ASTM D 4176 Colonial
Haze Scale
Several different types of measurement probes available to
fit all applications
Flow-through cells provide accurate measurements for very
low APHA or very high Saybolt ranges (optical paths up to
20 cm)
Typical applications are locations where automatic, on-line
color analysis of Diesel Fuel and Jet Fuel is required:
– fuel composition
Measurement probes are customizable and can be
engineered for specific application requirements.
Contact PSPI for more information.
CLEARVIEW COLORIMETER
42015 CLV
CLEARVIEW COLORIMETER
42015 CLV
S P E C I F I C A T I O N S
Performance
– Reproducibility: (dependent on installation)
Saybolt: ±1.3
Haze: ±0.2
ASTM: ±0.25
– Ambient Temperature: 10º to 50ºC (50 to 120ºF); weather protection
required; no direct sunlight
Sample Requirements
– Sample Pressure:
Insertion Probes: Up to 50 psig (3.5 bar g); Welker Insertion Device
recommended for higher